• PRODUCTOS
    • Monitorización
      • Connex® Spot Monitor
      • Connex® Vital Signs Monitor
      • Spot Vital Signs® 4400
    • Equipos RCP
      • Monitor RCP con desfibrilador
        • ZOLL AED 3®
        • ZOLL AED 3® BLS
        • ZOLL R Series®
      • Monitor de traslado RCP
        con desfibrilador
        • ZOLL X Series®
      • Ventiladores
        • ZOLL Z Vent®
      • Reanimador RCP
        • ZOLL AutoPulse®
    • Cardiología
      • Holter tensión arterial
        • Holter blood pressure ABPM 7100
      • Holter cardio
        • Digital recorder Holter H3+™
        • Digital recorder Holter H12+™
    • Diagnóstico
      • Otoscopios
        • Otoscope MacroView® Plus
        • 3.5V diagnostic otoscope
      • Oftalmoscopios
        • Ophthalmoscope PanOptic™ Plus
        • 3.5V Coaxial Ophthalmoscope
      • Esfigmomanómetros
        • Digital sphygmomanometer ProBP™ 2000
        • Connex® ProBP™ 3400 Digital Sphygmomanometer
      • Termometría
        • Thermometer SureTemp® Plus 692
        • Tympanic thermometer Braun® ThermoScan® PRO 6000
      • Accesorios
        • Rechargeable handle with 3.5 V Li-Ion battery
        • Single patient blood pressure cuffs FlexiPort® EcoCuff®
  • Equimed
  • Blog
  • Contact
© 2026 Equimed. Todos los derechos reservados.
Equimed – Medical equipment Equimed – Medical equipment
  • Products
    • monitorizacion-productos-header-menu

      Monitoring

      Connex® Spot Monitor
      Connex® Vital Signs Monitor
      Spot Vital Signs® 4400

      equipos-rcp-productos-header-menu

      CPR equipment

      CPR ​​Monitor with Defibrillator
      ZOLL AED 3®
      ZOLL AED 3® BLS
      ZOLL R Series®

      Defibrillator CPR Transport Monitor
      ZOLL X Series®
      ZOLL M2®

      Ventilators
      ZOLL Z Vent®

      CPR ​​Resuscitator
      ZOLL AutoPulse®

      cardiologia-productos-header-menu-02

      Cardiology

      Holter blood pressure
      ABPM 7100

      Holter cardio
      Holter H3+™
      Holter H12+™

      Electrocardiograph
      ELI™ 280

      diagnostico-productos-header-menu

      Diagnosis

      Otoscopes
      MacroView® Plus
      3.5V Diagnostic

      Ophthalmoscopes
      PanOptic™ Plus
      3.5V Autostep Coaxial

      Sphygmomanometers
      ProBP™ 2000
      Connex® ProBP™ 3400

      Thermometry
      SureTemp® Plus 692
      Braun® ThermoScan® PRO 6000

      Accessories
      3.5V Lithium ion battery handle
      FlexiPort® EcoCuff®
      Blood pressure cuffs 1 patient

      Gynecology
      KleenSpec Serie 590 Premium
      KleenSpec Serie 590 Premium
      with smoke evacuator

      esterilizacion-productos-header-menu

      Sterilization

      Rolls
      Flat Rolls
      Tyvek® Rolls

      Bags
      Flat pouches
      Organizer
      Self-sealing bags

      Control Systems
      Steam tape
      Chemical indicator strips
      EO Gas Indicator Strips
      Bowie & Dick Test
      Self-adhesive ETO Indicators

      Packaging solutions
      Crepe paper
      SMS paper

      Heat sealers
      Euroseal Infinity
      Euromatic® Plus

  • Equimed
  • News
  • Contact
  • Products
  • Equimed
  • News
  • Contact
  • Products
    • monitorizacion-productos-header-menu

      Monitoring

      Connex® Spot Monitor
      Connex® Vital Signs Monitor
      Spot Vital Signs® 4400

      equipos-rcp-productos-header-menu

      CPR equipment

      CPR ​​Monitor with Defibrillator
      ZOLL AED 3®
      ZOLL AED 3® BLS
      ZOLL R Series®

      Defibrillator CPR Transport Monitor
      ZOLL X Series®
      ZOLL M2®

      Ventilators
      ZOLL Z Vent®

      CPR ​​Resuscitator
      ZOLL AutoPulse®

      cardiologia-productos-header-menu-02

      Cardiology

      Holter blood pressure
      ABPM 7100

      Holter cardio
      Holter H3+™
      Holter H12+™

      Electrocardiograph
      ELI™ 280

      diagnostico-productos-header-menu

      Diagnosis

      Otoscopes
      MacroView® Plus
      3.5V Diagnostic

      Ophthalmoscopes
      PanOptic™ Plus
      3.5V Autostep Coaxial

      Sphygmomanometers
      ProBP™ 2000
      Connex® ProBP™ 3400

      Thermometry
      SureTemp® Plus 692
      Braun® ThermoScan® PRO 6000

      Accessories
      3.5V Lithium ion battery handle
      FlexiPort® EcoCuff®
      Blood pressure cuffs 1 patient

      Gynecology
      KleenSpec Serie 590 Premium
      KleenSpec Serie 590 Premium
      with smoke evacuator

      esterilizacion-productos-header-menu

      Sterilization

      Rolls
      Flat Rolls
      Tyvek® Rolls

      Bags
      Flat pouches
      Organizer
      Self-sealing bags

      Control Systems
      Steam tape
      Chemical indicator strips
      EO Gas Indicator Strips
      Bowie & Dick Test
      Self-adhesive ETO Indicators

      Packaging solutions
      Crepe paper
      SMS paper

      Heat sealers
      Euroseal Infinity
      Euromatic® Plus

  • Equimed
  • News
  • Contact
  • Products
  • Equimed
  • News
  • Contact
En
  • Catalan
  • Spanish
Equimed – Medical equipment
En
  • Catalan
  • Spanish
CPR equipment medical equipment

Defibrillator maintenance: keeping an AED ready between uses

Defibrillator maintenance should answer a specific question: is the AED ready now, in its assigned location and with the supplies it needs? In a healthcare facility or care home, answering that means checking the device, inspecting its consumables and knowing who will resolve a problem. A partially addressed alert leaves the question open.

Keeping an automated external defibrillator (AED) ready requires the manufacturer’s checks, a physical inspection, in-date consumables and clearly assigned responsibilities. Each issue needs an action and confirmation that it has been resolved. The record should show the device’s checked status, who inspected it and when it needs checking again.

This guide is for equipment managers, senior nursing staff, maintenance teams and procurement staff. It sets out a way to organise work for devices on standby or awaiting replacement supplies. During an emergency, follow the clinical protocol and the device’s instructions; administrative checks must not delay care.

AED readiness depends on both the device and its location

An AED may display a favourable status indicator but be hidden behind furniture. It may also be in its assigned place while replacement supplies are stored on another floor. These are hypothetical examples, but they show why two different checks matter: the status reported by the device and the conditions in which someone will find and use it.

The British Heart Foundation recommends a physical inspection in addition to self-tests. This helps identify signs of use or damage and check consumables and the battery. It also directs users to the manual for their model: automatic checks and alerts are not identical across AEDs. [2]

Two complementary levels of checking
AreaWhat to checkWhat must be resolved
Status reported by the AEDIndicator and messages defined by the manufacturer.Correct interpretation for that model; action on any alert.
Access to the AEDActual location, signage, obstacles and access during the intended hours.The device can be found and reached under the conditions stated.
Available suppliesPresence, condition and dates of the required materials.Replacement arranged before an item is missing or unusable.
Inspection arrangementsLast check, next review and assigned person.Continuity when shifts change or the usual owner is absent.

Proposed operational check distinguishing self-tests from physical inspection. The model’s manual defines the specific technical checks. [2, 3]

On the ZOLL AED 3 supplied by EQUIMED, a green status indicator means that the device passed its most recent self-test. Its manual says to check the status weekly or according to local guidance. [1] That check belongs in the facility’s routine, alongside inspection of the location where the AED is kept.

The indicator reports the device’s status. Checking the point of use completes the assessment.
ZOLL AED 3 in sharp focus with a hospital resuscitation bay softly blurred to the right
The AED 3 indicator provides information about the device; checking the point of use also establishes whether it can be reached.

The inspection record should identify the location precisely enough. “Reception” may be unambiguous in a small practice but insufficient in a building with several entrances. Recording the facility, floor and exact point helps prevent a correct entry from being attached to the wrong AED.

Defibrillator maintenance also depends on dates and lead times

Checking an expiry date is useful while there is still time to act. Procurement needs to know how long the correct replacement takes to arrive and who will order it. Without that information, an inspection can spot a problem early yet still end in avoidable downtime.

Keep a short record for each unit: AED identification, compatible consumable references, dates shown on their labels, where replacements are stored and the supplier contact. Installation dates, expiry dates and planned replacement dates must be distinguished; they are not interchangeable. When a replacement arrives, check the reference of the item delivered, not just the product family name.

For electrodes, the packaging is part of the check. The AED 3 manual says that the sealed pack must not be opened until use and that electrodes must not be reused. [1] Checking dates and references must not break that seal.

Prepare the replacement before it is needed.
ZOLL AED 3 and its electrode pack on a bright clinical surface
Identify electrodes by compatibility and reference; the pack should remain sealed until use.

A recorded date should allow time to arrange a replacement

As a planning rule, place the order according to the real supply lead time and the facility’s arrangements. This proposal sets no universal number of days or common service life for all batteries and electrodes. Follow the applicable labels and instructions.

Where several units are in service, sorting consumables by their next replacement date helps determine the next order. The facility’s existing management system may be enough; another application is useful only if it improves follow-up. What matters is updating the record for the correct AED after a replacement and identifying the material installed.

For a new purchase, ask for separate details of the device, consumables, support and any connectivity services. This makes it easier to assess continuity of supply and costs during use without confusing an included feature with an optional service. The quotation should reflect the configuration the facility will actually receive.

An issue needs a decision, not just an entry in the log

“Checked” is too broad when something remains unresolved. To organise the work, describe the finding and assign an action. The table suggests management decisions; it does not replace the manufacturer’s fault-finding instructions or authorise technical work outside a staff member’s competence.

From a finding to a specific action
FindingOperational decisionEvidence for closure
Access obstructed or location differs from the recordRestore access or correct the location and internal communication.Point checked and location record updated.
Replacement due soon; material still usableAssign the order, owner and follow-up date.Replacement received and change recorded when due.
Material missing, damaged or expiredArrange replacement and the agreed cover while the issue is resolved.Correct material available and applicable checks completed.
Device fault alertFollow the model’s instructions and escalate to qualified support.Action taken and final status documented.
Remote report not receivedCheck physical device status and communication channel separately.Status checked and connection reviewed; each issue closed in its own right.

Proposed management framework. The device’s technical status and suitability for return to service are determined according to its manual and the facility’s protocol.

A battery alert and an interrupted connection do not mean the same thing. Nor do they support the same conclusion about readiness. Keeping them separate prevents a network issue from being treated as an AED repair, or receipt of an email from being mistaken for confirmation that someone has inspected the device.

If a device must be removed from service, the organisation needs to know what alternative remains available. That arrangement should be agreed in advance according to the facility, its resources and the contracted service. Resuscitation Council UK advises clarifying temporary replacement arrangements at the purchasing stage; they should not be assumed to be included by every supplier. [3]

Wi-Fi supports monitoring when someone is ready to respond

ZOLL documents Wi-Fi connectivity options for remote status monitoring on the AED 3. [5] Before adding them to a facility’s workflow, confirm the device configuration, the service used and its conditions. The presence of Wi-Fi alone does not establish that the system is configured or that notifications reach the intended recipient.

An alert needs a recipient and a response.
Two healthcare professionals reviewing information beside a ZOLL AED 3
Monitoring works only when staff know who receives alerts and who confirms the action taken.

A useful setup test follows the whole route: the correct device is identified, the message arrives, it reaches the right person and a response is defined. One practical acceptance question is this: if an issue appears while the usual owner is away, who will see it and how will they know what to do?

For example, in a hypothetical situation, a facility changes its wireless network and reports stop arriving. The AED may need an on-site check while the communication problem is investigated. Restoring the network and checking the AED are related tasks, but each needs its own result; completing one does not prove that the other has been done.

Status monitoring should also be kept distinct from managing data from a clinical event. The tasks may share tools, but they serve different purposes. Access and responsibilities should match each task, without bringing clinical information into a maintenance record when it is not needed to resolve the issue.

After use, returning the AED to service needs a clear endpoint

The AED 3 manual covers inspecting and cleaning the device, replacing the electrodes and completing the relevant check before it is returned to service. [1] The facility’s process should show when those tasks are finished and who confirms the result.

  1. Identify the situation

    Record which unit was used, where it is and who will coordinate its recovery. If a temporary alternative is available, communicate it through the agreed channel.

  2. Complete the required work

    Replace supplies and carry out the checks in the manual. Cleaning must use products and methods permitted for the device and its accessories.

  3. Confirm the result

    Check the final status and the presence of the required supplies. An order sent to a supplier is not the same as material installed or a completed inspection.

  4. Restore and record

    Return the AED to its assigned point, update its status and tell those who need to know that it is available again.

Replace, check and confirm before closing the issue.
Healthcare professional placing a ZOLL AED 3 in the facility cabinet
Record the AED’s return to its cabinet after replacement supplies and the relevant checks have been completed.

The record should distinguish between “action requested”, “action completed” and “result confirmed”. This makes outstanding work visible without reconstructing an email trail. It also prevents an issue from being closed simply because the AED has physically returned to its cabinet.

Review of the clinical event follows its own process. Resuscitation Council UK addresses data retrieval and post-event review. [3] Both workstreams need coordinating so that someone remains responsible for the device throughout recovery, without assigning clinical functions to the technical maintenance record.

Who checks, who resolves and who confirms?

The MHRA’s guidance on managing medical devices stresses the need for clear responsibilities and for information to pass between users and maintenance staff. [4] The allocation below can be adapted to the size of the facility; it does not require new job titles. One person may carry out several functions if they have the training and resources to do so and cover is arranged for their absence.

Responsibilities and evidence
FunctionProposed responsibilityWhat to record
Check at the point of useCheck visible status, location and required supplies.Date, unit, result and findings.
Equipment coordinationAssign follow-up, deputy cover and an alternative if the AED is unavailable.Issue owner and next action.
Procurement or storesObtain the identified replacement and deliver it to the correct point.Reference, receipt and destination.
Qualified technical supportCarry out work requiring technical intervention.Work completed and results of checks.
Confirmation of closureVerify that the agreed conditions for return to service have been met.Final status, date and person confirming it.

Proposed allocation of work; adapt it to the facility’s skills, contracts and procedures.

A deputy must be able to continue the work from the information available. “Spoken to maintenance” does not say who was contacted, what is expected or when to check for a response. An action linked to a named person and a status can be followed up even after a shift change.

Four hospital stages: detect, assign, resolve and confirm AED readiness
Detect, assign, resolve and confirm links the device check to a documented return to service.

What to agree when introducing an AED

When the device is delivered, check how it will be kept ready. The list can guide the discussion between the facility and its supplier. It also connects with EQUIMED’s English-language overview of medical equipment maintenance within the equipment lifecycle.

  • Exact model and configuration, with the relevant documentation available to staff.
  • Compatible consumables, order references and replacement arrangements.
  • Assigned location, intended access and a named owner with deputy cover.
  • User checks and work reserved for qualified technical support.
  • Connectivity conditions, alert recipients and a test of message delivery, if that service is contracted.
  • Withdrawal procedure, temporary alternative and criteria for confirming return to service.

Ask for a demonstration of this workflow to assess how the device will be supported in practice. The discussion can end with a short, verifiable record: who checks the AED, how a finding is reported, who resolves it and what evidence allows the issue to be closed.

Frequently asked questions about AED checks

What does defibrillator maintenance include in a healthcare facility?

It includes the tasks specified by the manufacturer and the arrangements needed to keep the AED ready: inspections, consumables, issue management and recording the outcome. The precise scope depends on the model, its configuration and the facility’s procedures.

How often should an AED be checked?

Use the model’s manual and the guidance applicable to the facility to set the schedule. Distinguish checks by the responsible person from the device’s self-tests, and include checks needed after use or an issue.

Does a favourable status indicator replace a check at the point of use?

No. Check the device’s reported status alongside its location, access and supplies. An indicator does not show every change that may occur around the AED.

Does Wi-Fi connectivity guarantee that an AED is ready?

No. Remote communication can support monitoring when it is configured and working. Someone must receive alerts, respond to them and confirm that any required action has been completed.

When can an AED issue be closed?

Close it after the required action is complete and its outcome has been confirmed under the manual and the facility’s procedure. The record should identify the unit, the work done, the final status and the person confirming closure.

Agree the checking process alongside the equipment

If you are considering the ZOLL AED 3, contact EQUIMED to discuss configuration, compatible consumables and support arrangements. Bringing your access, monitoring and replacement requirements to that discussion will help define a proposal for the facility.

Sources and documentation

  1. ZOLL AED 3. Operator’s Guide, ref. 9650-003750-01, revision A, October 2023. Status indicator and checks, pp. 10–11; maintenance and return to service, pp. 19–24.
  2. British Heart Foundation. Maintaining your defibrillator, 6 November 2023. Physical inspection alongside self-tests and checking consumables.
  3. Resuscitation Council UK and British Heart Foundation. A guide to AEDs, June 2019. Sections 12–14: accessibility, responsibilities, maintenance and post-use recovery. UK organisational guidance; not presented as Spanish regulation.
  4. MHRA. Managing Medical Devices, January 2021. Sections 2, 5, 6 and 8: responsibilities, receipt, training, maintenance and records.
  5. ZOLL. ZOLL AED 3: features and connectivity. Remote status monitoring subject to optional connectivity and the applicable configuration.

AED checksAED self-testdefibrillator consumablesdevice readinesselectrode replacementZOLL AED 3

Latest news

Body temperature measurement: comparing readings between shifts

Body temperature measurement: comparing readings between shifts

A practical approach to recording temperature readings and preserving their context between shifts.
Defibrillator maintenance: keeping an AED ready between uses

Defibrillator maintenance: keeping an AED ready between uses

A practical guide to checking, restocking and returning an AED to service between uses.
Pulse oximetry: what to check when SpO₂ does not match the clinical picture

Pulse oximetry: what to check when SpO₂ does not match the clinical picture

Criteria for reviewing a discordant SpO₂ reading and selecting equipment and sensors for the intended clinical use.
Transport ventilator: what to check before, during, and after moving a patient

Transport ventilator: what to check before, during, and after moving a patient

A practical guide to checking a transport ventilator’s power, oxygen supply, circuit and alarms before patient transfer, monitoring the journey and handing over safely.
  • Pulse oximetry: what to check when SpO₂ does not match the clinical picture
    Previous PostPulse oximetry: what to check when SpO₂ does not match the clinical picture
  • Next PostBody temperature measurement: comparing readings between shifts
    Pulse oximetry: what to check when SpO₂ does not match the clinical picture
logo-equimed

An expert team in medical equipment

Sant Lluc, 69-81 08918 Badalona (Barcelona)
+34 933 992 058  equimed@equimed.es

© 2026 Equimed. All rights reserved.
Privacy Policy. Legal Notice. Cookies Policy. Web design NEXTPAINT

    Copy