This guide is for equipment managers, senior nursing staff, maintenance teams and procurement staff. It sets out a way to organise work for devices on standby or awaiting replacement supplies. During an emergency, follow the clinical protocol and the device’s instructions; administrative checks must not delay care.
AED readiness depends on both the device and its location
An AED may display a favourable status indicator but be hidden behind furniture. It may also be in its assigned place while replacement supplies are stored on another floor. These are hypothetical examples, but they show why two different checks matter: the status reported by the device and the conditions in which someone will find and use it.
The British Heart Foundation recommends a physical inspection in addition to self-tests. This helps identify signs of use or damage and check consumables and the battery. It also directs users to the manual for their model: automatic checks and alerts are not identical across AEDs. [2]
| Area | What to check | What must be resolved |
|---|---|---|
| Status reported by the AED | Indicator and messages defined by the manufacturer. | Correct interpretation for that model; action on any alert. |
| Access to the AED | Actual location, signage, obstacles and access during the intended hours. | The device can be found and reached under the conditions stated. |
| Available supplies | Presence, condition and dates of the required materials. | Replacement arranged before an item is missing or unusable. |
| Inspection arrangements | Last check, next review and assigned person. | Continuity when shifts change or the usual owner is absent. |
Proposed operational check distinguishing self-tests from physical inspection. The model’s manual defines the specific technical checks. [2, 3]
On the ZOLL AED 3 supplied by EQUIMED, a green status indicator means that the device passed its most recent self-test. Its manual says to check the status weekly or according to local guidance. [1] That check belongs in the facility’s routine, alongside inspection of the location where the AED is kept.

The inspection record should identify the location precisely enough. “Reception” may be unambiguous in a small practice but insufficient in a building with several entrances. Recording the facility, floor and exact point helps prevent a correct entry from being attached to the wrong AED.
Defibrillator maintenance also depends on dates and lead times
Checking an expiry date is useful while there is still time to act. Procurement needs to know how long the correct replacement takes to arrive and who will order it. Without that information, an inspection can spot a problem early yet still end in avoidable downtime.
Keep a short record for each unit: AED identification, compatible consumable references, dates shown on their labels, where replacements are stored and the supplier contact. Installation dates, expiry dates and planned replacement dates must be distinguished; they are not interchangeable. When a replacement arrives, check the reference of the item delivered, not just the product family name.
For electrodes, the packaging is part of the check. The AED 3 manual says that the sealed pack must not be opened until use and that electrodes must not be reused. [1] Checking dates and references must not break that seal.

Where several units are in service, sorting consumables by their next replacement date helps determine the next order. The facility’s existing management system may be enough; another application is useful only if it improves follow-up. What matters is updating the record for the correct AED after a replacement and identifying the material installed.
For a new purchase, ask for separate details of the device, consumables, support and any connectivity services. This makes it easier to assess continuity of supply and costs during use without confusing an included feature with an optional service. The quotation should reflect the configuration the facility will actually receive.
An issue needs a decision, not just an entry in the log
“Checked” is too broad when something remains unresolved. To organise the work, describe the finding and assign an action. The table suggests management decisions; it does not replace the manufacturer’s fault-finding instructions or authorise technical work outside a staff member’s competence.
| Finding | Operational decision | Evidence for closure |
|---|---|---|
| Access obstructed or location differs from the record | Restore access or correct the location and internal communication. | Point checked and location record updated. |
| Replacement due soon; material still usable | Assign the order, owner and follow-up date. | Replacement received and change recorded when due. |
| Material missing, damaged or expired | Arrange replacement and the agreed cover while the issue is resolved. | Correct material available and applicable checks completed. |
| Device fault alert | Follow the model’s instructions and escalate to qualified support. | Action taken and final status documented. |
| Remote report not received | Check physical device status and communication channel separately. | Status checked and connection reviewed; each issue closed in its own right. |
Proposed management framework. The device’s technical status and suitability for return to service are determined according to its manual and the facility’s protocol.
A battery alert and an interrupted connection do not mean the same thing. Nor do they support the same conclusion about readiness. Keeping them separate prevents a network issue from being treated as an AED repair, or receipt of an email from being mistaken for confirmation that someone has inspected the device.
If a device must be removed from service, the organisation needs to know what alternative remains available. That arrangement should be agreed in advance according to the facility, its resources and the contracted service. Resuscitation Council UK advises clarifying temporary replacement arrangements at the purchasing stage; they should not be assumed to be included by every supplier. [3]
Wi-Fi supports monitoring when someone is ready to respond
ZOLL documents Wi-Fi connectivity options for remote status monitoring on the AED 3. [5] Before adding them to a facility’s workflow, confirm the device configuration, the service used and its conditions. The presence of Wi-Fi alone does not establish that the system is configured or that notifications reach the intended recipient.

A useful setup test follows the whole route: the correct device is identified, the message arrives, it reaches the right person and a response is defined. One practical acceptance question is this: if an issue appears while the usual owner is away, who will see it and how will they know what to do?
For example, in a hypothetical situation, a facility changes its wireless network and reports stop arriving. The AED may need an on-site check while the communication problem is investigated. Restoring the network and checking the AED are related tasks, but each needs its own result; completing one does not prove that the other has been done.
Status monitoring should also be kept distinct from managing data from a clinical event. The tasks may share tools, but they serve different purposes. Access and responsibilities should match each task, without bringing clinical information into a maintenance record when it is not needed to resolve the issue.
After use, returning the AED to service needs a clear endpoint
The AED 3 manual covers inspecting and cleaning the device, replacing the electrodes and completing the relevant check before it is returned to service. [1] The facility’s process should show when those tasks are finished and who confirms the result.
Identify the situation
Record which unit was used, where it is and who will coordinate its recovery. If a temporary alternative is available, communicate it through the agreed channel.
Complete the required work
Replace supplies and carry out the checks in the manual. Cleaning must use products and methods permitted for the device and its accessories.
Confirm the result
Check the final status and the presence of the required supplies. An order sent to a supplier is not the same as material installed or a completed inspection.
Restore and record
Return the AED to its assigned point, update its status and tell those who need to know that it is available again.

The record should distinguish between “action requested”, “action completed” and “result confirmed”. This makes outstanding work visible without reconstructing an email trail. It also prevents an issue from being closed simply because the AED has physically returned to its cabinet.
Review of the clinical event follows its own process. Resuscitation Council UK addresses data retrieval and post-event review. [3] Both workstreams need coordinating so that someone remains responsible for the device throughout recovery, without assigning clinical functions to the technical maintenance record.
Who checks, who resolves and who confirms?
The MHRA’s guidance on managing medical devices stresses the need for clear responsibilities and for information to pass between users and maintenance staff. [4] The allocation below can be adapted to the size of the facility; it does not require new job titles. One person may carry out several functions if they have the training and resources to do so and cover is arranged for their absence.
| Function | Proposed responsibility | What to record |
|---|---|---|
| Check at the point of use | Check visible status, location and required supplies. | Date, unit, result and findings. |
| Equipment coordination | Assign follow-up, deputy cover and an alternative if the AED is unavailable. | Issue owner and next action. |
| Procurement or stores | Obtain the identified replacement and deliver it to the correct point. | Reference, receipt and destination. |
| Qualified technical support | Carry out work requiring technical intervention. | Work completed and results of checks. |
| Confirmation of closure | Verify that the agreed conditions for return to service have been met. | Final status, date and person confirming it. |
Proposed allocation of work; adapt it to the facility’s skills, contracts and procedures.
A deputy must be able to continue the work from the information available. “Spoken to maintenance” does not say who was contacted, what is expected or when to check for a response. An action linked to a named person and a status can be followed up even after a shift change.

What to agree when introducing an AED
When the device is delivered, check how it will be kept ready. The list can guide the discussion between the facility and its supplier. It also connects with EQUIMED’s English-language overview of medical equipment maintenance within the equipment lifecycle.
- Exact model and configuration, with the relevant documentation available to staff.
- Compatible consumables, order references and replacement arrangements.
- Assigned location, intended access and a named owner with deputy cover.
- User checks and work reserved for qualified technical support.
- Connectivity conditions, alert recipients and a test of message delivery, if that service is contracted.
- Withdrawal procedure, temporary alternative and criteria for confirming return to service.
Ask for a demonstration of this workflow to assess how the device will be supported in practice. The discussion can end with a short, verifiable record: who checks the AED, how a finding is reported, who resolves it and what evidence allows the issue to be closed.
Frequently asked questions about AED checks
What does defibrillator maintenance include in a healthcare facility?
It includes the tasks specified by the manufacturer and the arrangements needed to keep the AED ready: inspections, consumables, issue management and recording the outcome. The precise scope depends on the model, its configuration and the facility’s procedures.
How often should an AED be checked?
Use the model’s manual and the guidance applicable to the facility to set the schedule. Distinguish checks by the responsible person from the device’s self-tests, and include checks needed after use or an issue.
Does a favourable status indicator replace a check at the point of use?
No. Check the device’s reported status alongside its location, access and supplies. An indicator does not show every change that may occur around the AED.
Does Wi-Fi connectivity guarantee that an AED is ready?
No. Remote communication can support monitoring when it is configured and working. Someone must receive alerts, respond to them and confirm that any required action has been completed.
When can an AED issue be closed?
Close it after the required action is complete and its outcome has been confirmed under the manual and the facility’s procedure. The record should identify the unit, the work done, the final status and the person confirming closure.
Agree the checking process alongside the equipment
If you are considering the ZOLL AED 3, contact EQUIMED to discuss configuration, compatible consumables and support arrangements. Bringing your access, monitoring and replacement requirements to that discussion will help define a proposal for the facility.
Sources and documentation
- ZOLL AED 3. Operator’s Guide, ref. 9650-003750-01, revision A, October 2023. Status indicator and checks, pp. 10–11; maintenance and return to service, pp. 19–24.
- British Heart Foundation. Maintaining your defibrillator, 6 November 2023. Physical inspection alongside self-tests and checking consumables.
- Resuscitation Council UK and British Heart Foundation. A guide to AEDs, June 2019. Sections 12–14: accessibility, responsibilities, maintenance and post-use recovery. UK organisational guidance; not presented as Spanish regulation.
- MHRA. Managing Medical Devices, January 2021. Sections 2, 5, 6 and 8: responsibilities, receipt, training, maintenance and records.
- ZOLL. ZOLL AED 3: features and connectivity. Remote status monitoring subject to optional connectivity and the applicable configuration.