• PRODUCTOS
    • Monitorización
      • Connex® Spot Monitor
      • Connex® Vital Signs Monitor
      • Spot Vital Signs® 4400
    • Equipos RCP
      • Monitor RCP con desfibrilador
        • ZOLL AED 3®
        • ZOLL AED 3® BLS
        • ZOLL R Series®
      • Monitor de traslado RCP
        con desfibrilador
        • ZOLL X Series®
      • Ventiladores
        • ZOLL Z Vent®
      • Reanimador RCP
        • ZOLL AutoPulse®
    • Cardiología
      • Holter tensión arterial
        • Holter blood pressure ABPM 7100
      • Holter cardio
        • Digital recorder Holter H3+™
        • Digital recorder Holter H12+™
    • Diagnóstico
      • Otoscopios
        • Otoscope MacroView® Plus
        • 3.5V diagnostic otoscope
      • Oftalmoscopios
        • Ophthalmoscope PanOptic™ Plus
        • 3.5V Coaxial Ophthalmoscope
      • Esfigmomanómetros
        • Digital sphygmomanometer ProBP™ 2000
        • Connex® ProBP™ 3400 Digital Sphygmomanometer
      • Termometría
        • Thermometer SureTemp® Plus 692
        • Tympanic thermometer Braun® ThermoScan® PRO 6000
      • Accesorios
        • Rechargeable handle with 3.5 V Li-Ion battery
        • Single patient blood pressure cuffs FlexiPort® EcoCuff®
  • Equimed
  • Blog
  • Contact
© 2026 Equimed. Todos los derechos reservados.
Equimed – Medical equipment Equimed – Medical equipment
  • Products
    • monitorizacion-productos-header-menu

      Monitoring

      Connex® Spot Monitor
      Connex® Vital Signs Monitor
      Spot Vital Signs® 4400

      equipos-rcp-productos-header-menu

      CPR equipment

      CPR ​​Monitor with Defibrillator
      ZOLL AED 3®
      ZOLL AED 3® BLS
      ZOLL R Series®

      Defibrillator CPR Transport Monitor
      ZOLL X Series®
      ZOLL M2®

      Ventilators
      ZOLL Z Vent®

      CPR ​​Resuscitator
      ZOLL AutoPulse®

      cardiologia-productos-header-menu-02

      Cardiology

      Holter blood pressure
      ABPM 7100

      Holter cardio
      Holter H3+™
      Holter H12+™

      Electrocardiograph
      ELI™ 280

      diagnostico-productos-header-menu

      Diagnosis

      Otoscopes
      MacroView® Plus
      3.5V Diagnostic

      Ophthalmoscopes
      PanOptic™ Plus
      3.5V Autostep Coaxial

      Sphygmomanometers
      ProBP™ 2000
      Connex® ProBP™ 3400

      Thermometry
      SureTemp® Plus 692
      Braun® ThermoScan® PRO 6000

      Accessories
      3.5V Lithium ion battery handle
      FlexiPort® EcoCuff®
      Blood pressure cuffs 1 patient

      Gynecology
      KleenSpec Serie 590 Premium
      KleenSpec Serie 590 Premium
      with smoke evacuator

      esterilizacion-productos-header-menu

      Sterilization

      Rolls
      Flat Rolls
      Tyvek® Rolls

      Bags
      Flat pouches
      Organizer
      Self-sealing bags

      Control Systems
      Steam tape
      Chemical indicator strips
      EO Gas Indicator Strips
      Bowie & Dick Test
      Self-adhesive ETO Indicators

      Packaging solutions
      Crepe paper
      SMS paper

      Heat sealers
      Euroseal Infinity
      Euromatic® Plus

  • Equimed
  • News
  • Contact
  • Products
  • Equimed
  • News
  • Contact
  • Products
    • monitorizacion-productos-header-menu

      Monitoring

      Connex® Spot Monitor
      Connex® Vital Signs Monitor
      Spot Vital Signs® 4400

      equipos-rcp-productos-header-menu

      CPR equipment

      CPR ​​Monitor with Defibrillator
      ZOLL AED 3®
      ZOLL AED 3® BLS
      ZOLL R Series®

      Defibrillator CPR Transport Monitor
      ZOLL X Series®
      ZOLL M2®

      Ventilators
      ZOLL Z Vent®

      CPR ​​Resuscitator
      ZOLL AutoPulse®

      cardiologia-productos-header-menu-02

      Cardiology

      Holter blood pressure
      ABPM 7100

      Holter cardio
      Holter H3+™
      Holter H12+™

      Electrocardiograph
      ELI™ 280

      diagnostico-productos-header-menu

      Diagnosis

      Otoscopes
      MacroView® Plus
      3.5V Diagnostic

      Ophthalmoscopes
      PanOptic™ Plus
      3.5V Autostep Coaxial

      Sphygmomanometers
      ProBP™ 2000
      Connex® ProBP™ 3400

      Thermometry
      SureTemp® Plus 692
      Braun® ThermoScan® PRO 6000

      Accessories
      3.5V Lithium ion battery handle
      FlexiPort® EcoCuff®
      Blood pressure cuffs 1 patient

      Gynecology
      KleenSpec Serie 590 Premium
      KleenSpec Serie 590 Premium
      with smoke evacuator

      esterilizacion-productos-header-menu

      Sterilization

      Rolls
      Flat Rolls
      Tyvek® Rolls

      Bags
      Flat pouches
      Organizer
      Self-sealing bags

      Control Systems
      Steam tape
      Chemical indicator strips
      EO Gas Indicator Strips
      Bowie & Dick Test
      Self-adhesive ETO Indicators

      Packaging solutions
      Crepe paper
      SMS paper

      Heat sealers
      Euroseal Infinity
      Euromatic® Plus

  • Equimed
  • News
  • Contact
  • Products
  • Equimed
  • News
  • Contact
En
  • Spanish
  • Catalan
Equimed – Medical equipment
En
  • Spanish
  • Catalan
cardiology electrocardiographs medical equipment

ECG signal quality: how to identify and correct artefacts before repeating the recording

The first ECG shows an unstable baseline. It is repeated. The second recording contains the same noise and, when someone takes a closer look, the cause becomes apparent: an electrode that has lost adhesion, a cable pulling against the patient or someone who is uncomfortable and cannot relax their arms. The problem was not a shortage of recordings. It was repeating the ECG before finding out what was wrong.ECG signal quality is established long before anyone presses “Acquire”. It depends on patient and skin preparation, electrode condition, placement, cabling, the surrounding environment and the device settings. When one of these links fails, repeating the recording without a method usually reproduces the artefact and creates more work without adding useful information.

This guide offers a structured way to identify the cause, change one variable at a time and review the result before accepting the electrocardiogram. It is not intended to interpret the trace or replace local protocols, instructions for use or clinical judgement.

To improve ECG signal quality, check the following in order: patient comfort and stillness, skin preparation, electrode adhesion, freedom from cable tension, anatomical placement and electrocardiograph settings. If the artefact persists, change one variable and assess the effect before repeating the recording.

An artefact is not a clinical finding

In electrocardiography, an artefact is an alteration in the trace that does not originate from the heart’s actual electrical activity, but arises during signal acquisition or processing. It may appear as baseline drift, fine rapid noise, spikes, interruptions or a missing signal. Its appearance can guide the investigation, but rarely identifies a single cause on its own.

It is worth distinguishing terms that are often treated as interchangeable. Noise describes an unwanted component that contaminates the signal; interference points to a disturbance, usually from an external source; and an artefact is the alteration that ultimately appears in the recording. Not every technical error is an artefact either: incorrect electrode placement can change morphology without producing an obviously noisy trace.

This distinction matters. A clean signal does not mean that the ECG is normal, and a noisy signal does not prove that there is a cardiac abnormality. The first question is whether the recording is technically acceptable. A suitably qualified professional can then interpret it alongside the patient’s other clinical information.

For readers who first need to review the purpose of the test, Equimed explains what an electrocardiogram is and what it records. The focus here is narrower: preventing a technically poor acquisition from compromising the subsequent review.

What you seeWhat may be happeningFirst checkWhat to avoid
Baseline driftBreathing, movement or unstable contactPosture, comfort, adhesion and cable pullCompensating for it with filters alone
Fine high-frequency noiseMuscle tension or electrical interferenceRelaxation, proximity to power cables and mains filter settingAssuming that every channel has the same cause
Spikes or sudden jumpsMovement at an electrode, connection or cableIdentify the affected leads and follow their physical pathMoving every electrode at once
Flat or missing signalDisconnected electrode or leadCheck from the skin through to the acquisition moduleConfirming the recording without resolving the warning
Unexpected morphology without obvious noisePlacement error or possible lead reversalAnatomical landmarks and identification of each lead wireRepositioning by intuition without documenting it

The table is intended to help decide where to look first, not to diagnose a device fault or a clinical condition. If the problem continues after the approved procedure has been followed, the device or accessory should be managed in accordance with the organisation’s technical protocol.

Healthcare professional checking connections before acquiring an ECG with the ELI 280
A useful review begins before acquisition: patient, connections and signal are checked as parts of the same process.

ECG signal quality starts with the patient

In everyday practice, a short explanation can change the recording. Telling the patient what will happen, protecting their privacy and asking them to remain still and avoid speaking can make it easier to relax. The ELI® 280 user manual also emphasises that the patient should be lying down and comfortable before acquisition begins.

Some situations need a little more time: cold, tremor, pain, anxiety, breathing difficulty or a poorly supported position. Their effect should not simply be hidden on the screen. They should be addressed within the clinical limits of the case, the environment adapted where possible and anything that may affect the recording documented.

Skin and electrode form a single interface

Electrode-to-skin contact needs to remain stable. Hair, skin oils, moisture, product residue, dry skin or deteriorated adhesive may increase impedance or cause the contact to change with movement. Correct preparation is not a universal recipe: it depends on the electrode, its instructions for use, the condition of the skin and the approved protocol.

It is therefore unwise to improvise with products or techniques that have “always worked”. Recommendations may vary between consumables and institutions. For frail or older patients, and for anyone with vulnerable skin, preparation requires particular care to avoid discomfort or injury.

Skin preparation and electrode placement to improve ECG signal quality
Preparation should follow the electrode instructions, the organisation’s protocol and the characteristics of the patient’s skin.
LinkWhat to checkEvidence that it is readyWarning sign
PatientInformation, comfort, position and stillnessCan remain relaxed during acquisitionTremor, conversation, muscle tension or strained posture
SkinCleaning, dryness and approved preparationSurface suitable for stable contactMoisture, cream, hair or irritation
ElectrodeIntegrity, storage, gel and adhesionDoes not move and maintains contactOpen packaging, dryness, damage or a lifting edge
CableConnection, support, routing and lack of tensionMovement of the cable set does not pull on the electrodePulling, snagging, loose connector or visible damage
PlacementAnatomical landmarks and lead correspondencePosition consistent with the standardised methodApproximate positioning or reversal
DeviceIdentity, gain, speed, filters and warningsProtocol-compliant settings and a reviewable signalUnknown settings or unresolved warnings

How to identify the cause without repeating blindly

When a technical alteration appears, changing everything at once makes it impossible to know what solved it. A short sequence and one change at a time are more useful. If correcting the contact of one electrode removes the problem, the team has learnt something. If ten electrodes are repositioned, the filter changed and the cable replaced at the same time, the second ECG may be clean, but the healthcare team will not know why.

  1. Stop the recording from being accepted. Do not confirm or send it while a relevant technical doubt remains.
  2. Assess the extent. Determine whether the artefact appears in one lead, a particular group or all leads.
  3. Return to the patient. Check position, movement, speech, tremor and comfort.
  4. Follow the physical path. Review the skin, electrode, connector and cable for the affected leads.
  5. Confirm placement. Use anatomical landmarks rather than the position used for a previous ECG or a visual approximation.
  6. Review the settings. Check speed, gain and filters against the protocol. Do not use them to disguise poor acquisition.
  7. Acquire again and compare. Assess whether the change corrected the pattern before moving to another possible cause.
ELI 280 with ECG cables organised and supported to prevent tension on the electrodes
A supported cable set, free from pulling and snagging, prevents cable movement from being transferred to the contact point.

The affected leads can help focus the investigation

The trace does more than show that an alteration is present; it can sometimes narrow down where to look. The ELI® 280 manual relates certain combinations of affected limb leads to the electrode or limb that should be checked. This is a practical troubleshooting aid, not a rule for clinical interpretation.

Leads with artefact or abnormal signalFirst area to reviewStructured action
I and IIRA electrode or right arm movementCheck relaxation, contact, connector and RA cable
II and IIILL electrode or left leg movementCheck support, contact, connector and LL cable
I and IIILA electrode or left arm movementCheck relaxation, contact, connector and LA cable
One or more precordial leadsCorresponding electrode, preparation, connection or positionFollow each lead from the chest to the module
High-frequency noise in all leadsElectrical environment and configurationCheck power cables and the mains filter frequency

The value of this approach lies in reducing the field of investigation. If the alteration is concentrated in leads I and II, there is little point in starting by replacing every precordial electrode. Begin with what those leads have in common and then assess the effect of the correction.

A filter can clean up the appearance and alter the signal

Filters are necessary for managing unwanted signal components, but they are not a harmless eraser. The AHA, ACC and HRS scientific statement on ECG standardisation explains that certain filters can reduce baseline drift or high-frequency noise while also altering the fidelity of parts of the trace. The setting used should therefore be visible, understood and consistent with the procedure.

If the recording only appears acceptable after changing filters without correcting the physical cause, the acquisition still deserves further review.

The ELI® 280 allows the mains interference filter to be configured at 50 or 60 Hz and offers 40, 150 and 300 Hz print filters. The manual explains that these selections attenuate different frequencies and affect the fidelity of the displayed and printed trace. There is therefore no single setting that should be applied automatically in every situation. The organisation should define it for the intended use and retain that information with the recording.

How the ELI® 280 can support quality control

Preparation and placement remain human tasks, but the electrocardiograph can make certain problems visible before the examination is accepted. Equimed’s page for the ELI® 280 resting electrocardiograph documents simultaneous display of all twelve leads, connection confirmation, lead-fail indication and detection of possible lead reversals.

The device also includes Best 10. When the function is active, it analyses segments held in a five-minute buffer and selects ten seconds on the basis of high- and low-frequency noise. The manual also states that Best 10 is unavailable while certain lead-fail conditions remain. The sequence makes sense: resolve the connection first, then select the segment.

Best 10 can support review and reduce avoidable repeats, but it cannot turn incorrect placement into correct placement or replace professional assessment. It is one aid within a wider quality chain. Equimed’s guide to choosing a 12-lead electrocardiograph explores how these functions fit into clinical workflow, connectivity and organisational support.

Healthcare professional reviewing the quality of a printed trace and the ELI 280 screen
Reviewing before confirmation allows a technical problem to be corrected before the recording reaches the clinical archive.
Isometric infographic showing the route from an artefact to a usable ECG signal
Conceptual framework: correct the patient, skin, electrode, cable, placement and settings before repeating, reviewing or documenting.

Reducing repeat recordings takes more than training one person

When the same type of artefact appears repeatedly across different departments or shifts, the problem may not be an individual mistake. Electrodes may be stored poorly, cables replaced too late, settings left unstandardised or preparation performed differently by each professional. At that point, the question changes from “Who recorded this ECG?” to “Which part of the system is creating the variation?”

AreaSpecific responsibilityUseful evidenceProblem it helps prevent
Clinical teamPreparation, placement, review and documentationShared checklist and acceptance criteriaVariation between professionals
TrainingPractise with clean recordings and technical problemsObserved practice and periodic refresher trainingAutomatic repetition without identifying the cause
Procurement and storesControl references, storage and stock rotationBatches, expiry dates and storage conditionsDry or incompatible electrodes
Clinical engineeringReview cables, modules, battery and maintenanceHistory of incidents and replacementsIntermittent accessory failures
Clinical leadDefine settings and acceptance criteriaCurrent protocol and change recordInconsistent use of filters and settings
Quality teamAnalyse repeat recordings and recurring causesRecord sampling and corrective actionsTreating every artefact as an isolated event

Checklist before accepting the recording

  • The patient identity and request match.
  • The patient is informed, comfortable and as still as their condition allows.
  • The skin has been prepared in accordance with the protocol.
  • The electrodes are intact, adherent and within their conditions of use.
  • The cables are connected, supported and free from tension.
  • The leads follow the intended anatomical landmarks.
  • Speed, gain and filters are authorised for the examination.
  • All twelve leads are present and warnings have been resolved.
  • Relevant changes have been documented.
  • The recording has been reviewed before transmission or archiving.

A good signal is achieved by correcting, not simply repeating

The most reliable way to reduce artefacts is not to pursue a perfect-looking screen, but to make acquisition reproducible. Patient, skin, electrodes, cabling, placement and settings should be reviewed in the same order, even when the department is under pressure.

This method saves repeat recordings, but offers something more valuable: it makes it possible to explain why a recording was accepted, what was changed and what should be reviewed if the problem appears again. Technology helps professionals see and manage the signal. Quality emerges when that support is integrated into a procedure that people can apply and verify.

Frequently asked questions about ECG signal quality

What are the most common causes of artefacts in an ECG?

Common causes include movement, muscle tension, insufficient skin preparation, deteriorated electrodes, unstable contact, cable pull and electrical interference. Incorrect placement is another important technical error, although it does not always produce a visible artefact.

Should every noisy ECG be repeated?

Not automatically. The organisation’s acceptance criteria should be applied first, followed by identification and correction of the likely cause. Repeating the recording without changing anything may reproduce the same artefact.

Can filters correct poor acquisition?

No. Filters attenuate particular components and may alter the appearance and fidelity of the trace. They should be used according to protocol, not as a substitute for correct preparation, contact or placement.

Why does cable tension affect the ECG signal?

Movement or the weight of the cable can be transferred to the electrode and alter its contact with the skin. Supporting the cable set and avoiding pulling helps maintain a stable interface.

Can electrode positions be changed to reduce noise?

Only when justified by the clinical situation or protocol. The change should be documented because an alternative position can alter morphology and affect comparison with other recordings.

How does the ELI® 280 help control ECG signal quality?

The ELI® 280 displays all twelve leads, reports connection failures, can detect possible reversals and offers Best 10 to select a segment with less noise. These functions support, but do not replace, professional preparation and review.

Do you need to review your organisation’s ECG acquisition process?

Equimed can provide further documented information about the ELI® 280, review technical requirements and prepare a demonstration focused on acquisition, signal quality and workflow.

Request technical advice

Technical sources consulted

  • Baxter / Welch Allyn: ELI® 280 electrocardiograph user manual.
  • Equimed: ELI® 280 resting electrocardiograph.
  • ELI® 280 technical data sheet available from Equimed.
  • AHA/ACC/HRS: recommendations for the standardisation and interpretation of the electrocardiogram, part I.
  • NHS Supply Chain: clinical review of adult ECG electrodes.

12-lead ECGECG acquisitionECG artefactsECG signal qualityelèctrodesELI 280

Latest news

ECG signal quality: how to identify and correct artefacts before repeating the recording

ECG signal quality: how to identify and correct artefacts before repeating the recording

A practical guide to identifying artefacts in an ECG, correcting their source methodically and avoiding repeat recordings that do not produce a technically better signal.
How to Choose a 12-Lead Electrocardiograph for a Healthcare Facility

How to Choose a 12-Lead Electrocardiograph for a Healthcare Facility

A practical guide to choosing a 12-lead electrocardiograph by evaluating signal quality, clinical workflow, connectivity, security, lifecycle costs and technical support.
Medical equipment for Primary Care: what a consultation room needs in 2026

Medical equipment for Primary Care: what a consultation room needs in 2026

Discover what medical equipment a Primary Care consultation room needs in 2026 to cover vital signs, diagnostics, cardiology, connectivity, accessories and maintenance.
EUDAMED and medical equipment traceability: key considerations for hospitals in 2026

EUDAMED and medical equipment traceability: key considerations for hospitals in 2026

Medical equipment traceability will be a key priority in 2026 with EUDAMED and UDI.
  • How to Choose a 12-Lead Electrocardiograph for a Healthcare Facility
    Previous PostHow to Choose a 12-Lead Electrocardiograph for a Healthcare Facility
logo-equimed

An expert team in medical equipment

Sant Lluc, 69-81 08918 Badalona (Barcelona)
+34 933 992 058  equimed@equimed.es

© 2026 Equimed. All rights reserved.
Privacy Policy. Legal Notice. Cookies Policy. Web design NEXTPAINT

    Copy